管家婆正版资料网站

Sina Weibo | Tencent Weibo | Favorites Jiebo | Online Message | Site Map

Welcome to Wuxi Jiebo Instrument Technology Co., Ltd.!

杰博科技

Precision from instruments · Trust in Geber Precision instruments from

National Consulting Hotline: 400-858-0166

杰博科技

首页 » 杰博新闻中心 » 杰博资讯 » 行业资讯 » 直读光谱仪使用中高低标的作用及如何选择高低标 You are here: Home » Jiebo News Center » Jiebo News » Industry News » The role of high and low standards in the use of direct reading spectrometers and how to choose high and low standards

The role of high and low standards for direct reading spectrometers and how to choose high and low standards

】 Article Source: Editor-in-Chief: Release time: 2019-09-29 10:28:00 Hits: - [Big Middle Small]
In the process of using the direct-reading spectrometer , in addition to the elements of the sample itself, the vacuum, electronic circuits, ambient temperature, ambient humidity, constant temperature of the optical room, pollution of the optical path, and misalignment of the slit will cause the working curve to drift. As a result, measurement errors increase. In addition, the lens is contaminated to form a coating, which causes the electrode to be sharp during the excitation process, which increases the background of the curve display, and changes in argon flow, pressure, and purity can also cause the curve to drift. In particular, the spectral lines of the C, P, and S elements are located in the far ultraviolet region, and the changes are particularly sensitive, so it is very important to correct the working curve. Selecting a set of high and low standards and performing data calibration before analysis and calculation will save a lot of calibration time and simplify operation procedures.

image002 (6)

Normally samples need to be standardized before analysis. The calculation of standardized data is performed automatically by a computer. Selecting a standardized set of samples is difficult, including high and low levels of all elements to be analyzed. The direct-reading spectrometer standardized sample must excite the standardized sample while making the elemental working curve to ensure that the curve is not shifted. In the 1-point standardization, only high or low content standards are sufficient. If the sample can cover a wide range of content, and high accuracy is required at low content, 2-point standardization is required. At this time, two kinds of standard samples are needed, one for high standard and one for low standard. Or each standard can contain some high-concentration elements and other low-concentration elements, as long as they can include all elements. The high and low concentration values are not necessarily exactly at the highest or lowest end of the calibration range. But they should be close to the highest and lowest ends of the range, and there should be enough difference between them. The high and low standards are used to correct the deviation of the measured value from the working curve due to the change of the instrument over time, and then recalibrate to correct the drift of the instrument in the medium and long term.
In the spectral analysis, in order to directly use the original working curve, it is necessary to periodically recalibrate the original analysis curve with recalibration samples. When two-point recalibration is used, the element content of the two recalibration samples used should be sufficiently different, and ensured at the high and low ends of the elemental analysis range. In fact, the low-end recalibration samples often use "pure" matrix samples, which means that they actually contain only this alloy matrix, and the content of other elements should be zero, trace, or very small. Make a low. Theoretically, such two samples are sufficient to restandardize the instrument. However, in actual work, in order to ensure the accuracy of the instrument, it is generally impossible to cover all high-end and low-end of all elements with only two recalibration samples. Therefore, a larger number of samples are often required for recalibration. On the other hand, the recalibration procedure can be used for more than one alloy, and in principle it can be used for a group of alloys. At this time, multi-point recalibration is more convenient and effective.
行业资讯 Keywords in this article: Industry information

Wuxi Jiebo Instrument Technology Co., Ltd. All rights reserved Record number: 苏 ICP 备 10025668 号 -6
Phone: 400-858-0166 Fax: 0510-84601616
Address: 26-9 Huibei Road, Optoelectronic New Material Industrial Park, Liangxi District, Wuxi City
Technical Support: Avatar Networks

微信二维码 Wechat QR code

手机二维码 Scan your phone

管家婆资料免费公开 管家婆图香港正版 管家婆彩图大全资料 管家婆免费版资料大全 正版管家婆马报彩图 管家婆港彩免费资料大全 管家婆今期码报资料 管家婆资料大全管家婆2020 管家婆精选高手资料 香港管家婆马报彩图 管家婆资料大全 管家婆码报彩图马报料 管家婆期期准免费资料2020 管家婆马报图今晚资料143期 采金网免费资料管家婆 管家婆2020开奖结果 2020管家婆手机论坛 管家婆资料大全管家婆图片 管家婆资料大全管家婆一图 管家婆精选心水资料站